VENTICELL IL
ISO Class 5 & 7 Depyrogenation Ovens for Laboratories, Pharmacy and Industry
VENTICELL IL
Sterilization and Depyrogenation in Laboratories, Pharmacy and Industry
VENTICELL ® IL, ISO Class 5 and 7 ovens are designed for dry heat sterilization and depyrogenation of glassware and stainless at a temperature up to 280 °C.
The ISO class 5 or 7 is performed during the entire cycle, including heating and cooling. The chamber sizes can be customized, minimum size 700 liters. The ovens can be used in pharmaceutical manufacturing, laboratories and research.
Key Features
- Modular system allows variable individual device variation
- One-door and passthrough models
- Control by means of an industrial PLC system
- Sterilization chamber, doors, device frame and jacket are made of stainless steel for easy maintenance and long service lifetime
- Double semi-automatic door lock for maximal process safety
Technical Data
General Information
- Modular system allows variable individual device variation
- One-door and passthrough models
- Sterilization chamber, doors, device frame and jacket are made of stainless steel for easy maintenance and long service lifetime
- Double semi-automatic door lock for maximal process safety/li>
- Horizontal air flow in the chamber, powerful heating elements and highly efficient device insulation ensure short work process times and reduction of operating costs
- Control by means of an industrial PLC system
- Sterilization phase checking and recording both in graphic and numeric form in the course of the whole process
- Simple, intuitive device control by means of a touch panel with possibility of user modification of process parameters
- Various possibilities of the batch documentation processing
- Regulation of pressure inside the chamber by means of air pressure sensors in relation to sterile or non-sterile space
- Main doubled temperature sensors for independent work process control
- Auxiliary doubled temperature sensors for better process control
- Effective use of inner sterilization space
- Transport and loading system guarantees easy handling of the sterilized material by the operator
- Wide offer of optional equipment and accessories according to individual needs
Use
VENTICELL® IL is intended for thermally resistant, inflammable materials, e.g.:
- Empty glass products – glasses, ampoules, vials, bottles, vessels
- Metal materials in pharmaceutical industry – trays, containers, accessories and device parts
- Thermally stable basic pharmaceutical products and chemical substance
Sterilization & Depyrogenation
Sterilization is a process that ensures killing of all viable microorganisms, including spores, and which leads to an irreversible inactivation and killing of health-relevant worms and their eggs
During the sterilization cycle, the number of Bacillus subtilis microorganisms must be reduced by at least 6 orders of magnitude. The sterilizing effect in hot air sterilizers is achieved by heating the sterilized material to high temperatures .
Depyrogenation is a procedure reducing the number of bacterial endotoxins (pyrogens) by at least three orders by high temperature (250-280°C) acting for a given period of time.
Important parameters of the processes are the following ones:
- Accurate profile reached by a sophisticated heated air flow system and perfect construction design of the sterilization chamber
- Quick temperature rise and cooling
- Compliance with the regulations on clean rooms
GET IN TOUCH